May, 2002
Use this software screen to:
Display statistic calculation results:
1. Select [Auxiliary], [Data Statistics].
2. Use the pull-down menu by the index field to select an index date and time.
3. Specify a type of sample for statistic calculation.
Using your mouse, select the check box
of your choice: serum, urine, or other; routine, emergency, or STAT.
4. Move the cursor to the “Sample No.” column and enter a sample number range
. A
sample ID range can also be entered. For sample types that do not need to be specified,
enter a space. To specify the entire range of samples enter an "*."
5. After calculation, the results can be displayed in a list. Press function key F5 (data
display). The following data will be displayed in a list:
Specified index date and time
•
Sample type of the selected sample(s)
•
Number of samples selected
•
Analysis test name
•
Number of samples, mean, standard deviation, coefficient of variation (CV), range,
•
maximum value, and minimum value of each analysis test
6. Press function key F3 (print).
A window to start printing is displayed
.
7. Select “OK.”
Printing begins.
8. Optional Step: To add a search key, select function key F6 (search add). Set "Sex," "Age
(Low and high limit)," and 6 types of demographics.Select "Yes." The settings will be
saved as search keys.
Note
An asterisk (*) can only be used if specific demographic fields are used in sample processing.