D-102
Chapter D - Software
Automated Chemistry Analyzer
AU2700 User's Guide
May, 2002
Check range:
Set an upper limit and a lower limit of the normal value range of the inter-test
check. Specify a numerical value within the range between -9999999 and 9999999. Decimal
points can be used.
Decimal Places:
Displays the upper and lower limit of the range set in the "Decimal Places"
dialog box.
Use this software screen to:
Set inter-test check parameters:
Set the check name and check range for each inter-test check.
1. Select [Parameters], [Inter-Related Tests], [Checked Tests].
2. Enter an "Inter-Test Check Name."
Use the pull-down menu located by the "Inter-Test
Check Name" field. The </> buttons can also be used to select a test name.
3. Select a sample type.
Use the pull-down menu located by the "Type" field.
4. Press function key F4 (set). Set the following: Check name, Test name, Constant,
Formula, and Check range.
5. After the setting is complete, press function key F2 (exit).
Parameters will be saved.
Specify another inter-test check:
In the [Calculated Tests] screen, select function key F8 (Inter-test Check Number).
A window
appears.
Enter an inter-test check number between 1 and 10, then select "OK."
To change the
sample type, use the pull-down menu next to the "Type" field or use the </> buttons.