D-158
Chapter D - Software
Automated Chemistry Analyzer
AU2700 User's Guide
May, 2002
4.1 Data Statistics
F5
F6
F3
OLP3052
Specifying a search
key for samples
subject of statistic
processing
Referring to the
statistic calculation
results
(A)Auxiliary
(S)Data Statistics
Setting an additional
search key
Outputting the
statistic calculation
results for a data list
One of the tabs
can be selected
Button function
Described in
another section.
Press this key.
Main operation flow
Operation flow
1. 11/30/2000 12:30 - 1. 11/30/2000 12:30
ALARMCLEAR
?
STANDBY
Help
Exit
Data Statistics
11/30/2000 10:10
Index:
Sample No.
Sample ID
Routine
Serum
Emergency
STAT
( nnnn - nnnn)
( Ennn - Ennn)
( Pnnn - Pnnn)
Routine
Urine
Emergency
STAT
( nnnn - nnnn)
(UEnnn - UEnnn)
Routine
Others
Emergency
STAT
( nnnn - nnnn)
(XEnnn - XEnnn)
(UPnnn - UPnnn)
(XPnnn - XPnnn)
Data
Display
Search
Add
OLP1dats
Description:
This screen allows selection of the samples to be used in generating sample statistics. Statistics
include the number of samples, mean, SD, CV, range, and min./max. value. Statistics can be
calculated for routine, emergency, STAT, serum, urine, or other samples.
Left Index Box:
Select a starting range for review by using the pull-down menu.
Right Index Box:
Select an ending range for review by using the pull-down menu.
Serum, Urine, Other, Routine, Emergency, STAT:
Select the sample type by clicking on the
check box next to the type.
Sample No.:
Set the sample number range within the numbers printed in parenthesis. To
specify all samples, enter an asterisk (*). To leave the field blank, enter a space.
Sample ID:
Set the sample ID range within the range specified in [Parameters], [Requisition
Format]. To specify all samples IDs, enter an asterisk (*). To leave the field blank, enter a
space.
Содержание AU2700
Страница 17: ...A 4b Automated Chemistry Analyzer AU2700 User s Guide November 1 2009 This page intentionally left blank ...
Страница 178: ...D 4 Chapter D Software Automated Chemistry Analyzer AU2700 User s Guide March 31 2005 ...
Страница 260: ...D 86 Chapter D Software Automated Chemistry Analyzer AU2700 User s Guide May 2002 ...