Automated Chemistry Analyzer
AU2700 User's Guide
Chapter D - Software
D-193
Automated Chemistry Analyzer
AU2700 User's Guide
Chapter D - Software
D-193
March 20, 2009
4 . View the printout.
Analyze the calibration data printout:
If acceptable ranges are exceeded, an asterisk (*) will appear on the right of the data.
ISE CALIBRATION PRINT TIME [11-28-2000 09:43]
<<SERUM>>
MID STD. H MID STD. L MID
SLOPE FACTOR
Na 59.5 0.965 -37.9 -81.6 -91.9 -82.0
K 55.2 0.997 133.5 35.6 -0.8 30.7
Cl -49.9 0.947 1452.9 1504.2 1533.6 1510.2
* Voltage (MID)
FORMER LATTER The acceptable range between the former and latter is +/- 10
Na -83.4 -84.1
K 31.3 30.2
Cl 1507.1 1506.9
*VOLTAGE (STD. H)
1ST 2ND 3RD 4TH
The acceptable range between the 3rd and 4th value is +/- 5 for all electrolytes except urine K,
which is +/- 10
Na STD. H -38.6 -38.2 -37.9 -37.8
MID -82.4 -81.9 -81.7 -81.5
K STD. H 133.0 133.3 133.5 133.4
MID 35.0 35.6 35.8 35.4
C1 STD. H 1452.8 1452.5 1452.8 1452.9
MID 1504.0 1504.2 1504.0 1504.4
*VOLTAGE (STD. L)
1ST 2ND 3RD 4TH
Na STD. L -94.8 -92.1 -91.8 -92.0
MID -83.2 -83.1 -82.5 -81.5
K STD. L -2.6 -0.7 -0.6 -0.9
MID 30.0 29.9 30.3 31.0
C1 STD. L 1532.7 1532.3 1533.1 1534.1
MID 1508.6 1509.1 1509.7 1510.7
Perform a selectivity check from diagnostics:
1. Place the analyzer in a stop mode.
2. Select [Maintenance], [Maker Maintenance], [ISE Diag].
3. Select selectivity.
Start Check? Yes
See Also
For information on setting up a selectivity check, refer to the "ISE Maintenance" section located
in the Maintenance Chapter.
4. Verify the data printout results are within the acceptable range.