C-84
Chapter C - Basic Operations
Automated Chemistry Analyzer
AU2700 User's Guide
May, 2002
4.4 Editing Results
Analysis results can be directly modified and edited as necessary.
Caution
Editing results is an operation performed by the operator to forcibly change test results that were
analyzed according to a certain calibration and quality control. Follow specific laboratory protocols and
procedures for editing results.
Editing Results
1. On the main screen, select [Routine], [Data Management] and [Data Edit] in that order.
The screen for editing the data is displayed.
OLP1023
1989/07/07 12:30
ALARMCLEAR
?
STANDBY
Help
Exit
Test No.
Data Edit
1997/09/03 10:10
Index:
Test Name.:
Sample No.
Sample ID
Routine
Serum
Emergency
STAT
( nnnn - nnnn)
( Ennn - Ennn)
( Pnnn - Pnnn)
Routine
Urine
Emergency
STAT
(Unnn - Unnn)
(UEnnn - UEnnn)
Routine
Others
Emergency
STAT
(Xnnn - Xnnn)
(XEnnn - XEnnn)
(UPnnn - UPnnn)
(XPnnn - XPnnn)
Data
Display
Search
Add
2. Use the pull-down menus to select an index and test name.
3. Specify the type of the sample to be edited by clicking on the check box with the mouse.
4. Enter the range of sample numbers. It is also possible to enter the range of sample IDs.
Note:
To edit according to test number, press function key F8 (test no.). A dialog box
displays. Enter the test number and click the “OK” button.
5. Press function key F5 (data display).
The analysis result data displays for tests or
samples.