Automated Chemistry Analyzer
AU2700 User's Guide
Chapter D - Software
D-67
May, 2002
5. Set the display to either manual or auto.
If “Manual” is selected, also specify a start
data number and the number of data samples. The specified portion of the graph will be
enlarged.
When “Auto” is selected, the graph will display at the largest size at which all
the data can be properly represented.
Note
The X and Y axis indicate the following: X axis indicates the sample number. The Y axis (left side)
ind3SD, +2SD, +1SD, Mean, -1SD, -2SD, -3SD. The Y axis (right side) indicates each value of
the Y axis (left side)
Display previous results on each control sample:
1. Select [Routine], [QC Monitor], [Daily Control].
2. Select the "QC Monitor" tab.
3. Select the tests displayed on screen using your mouse.
QC Monitor Column displays:
*
Green
within ±1SD
1
Yellow, reverse video
Outside ±1SD and within ±2SD
2
Yellow, reverse video
Outside ±2SD and within ±3SD
3
Yellow, reverse video
Outside ±3SD
E
Red, normal video
Error, no data value
4. Press function key F6 (daily data).
The chart will display.
5. Press function key F2 (exit) twice.
Print the daily variation data:
1. Select [Routine], [QC Monitor], [Daily Control].
2. Select the tests displayed on screen using your mouse.
3. Press function key F6 (daily data).
4. Press function key F3 (print).
A window to specify the printing mode is displayed.
s
elect
one
of
the
followinG
printinG
modes
:
Statistic:
Prints only the statistical values.
Sta Data:
Prints the statistical values and numerical data.
Graph:
Prints only a graph. Comments will not be displayed.
Graph +
Comment:
Prints a graph and comment.
All Data +
Comment:
Prints the numerical data and comment.
5. Select “OK.”
Printing Begins
6. Press function key F2 (exit) twice.