ALARMCLEAR
?
1997/09/03 10:10
STANDBY
OLP1020
Errors found as below.Start Measure?
Yes
Start Routine S.No.:
Operator Name:
No
Serum
Urine
Others
Change Select
Check Start No
(P) Parameters
(A) Auxiliary
(M)Maintenance
(L) Logout
(R) Routine
(U) User
Tips
To change a setting such as the “Start Routine S. No.,” press function key F4 (set). Use the mouse to
select the field.
6. Check the start routine and urine S. No., operator name, and error list.
If a fatal error
(red) occurs, it is not possible to start analysis. It is possible to start analysis with
nonfatal (yellow) errors. A message describing the error is displayed. Check the
description of the error and press Yes to bypass the error and start analysis or press No
to perform corrective actions before analysis. Follow the necessary corrective actions to
recover from the error.
See Also
For information on recovering from errors, refer to the Error Flags and Troubleshooting Chapters.
Tip
An alarm will occur if: 1. A different reagent lot number is used from those used for reagent blank or
calibration. 2. If different R1 and R2 reagent lot numbers are mixed for the same analysis item. If a new
lot number is detected, perform a reagent blank or calibration on that test. The same reagent lot number
should be used for reagent blank, calibration and patient samples.
If these problems are not corrected,
accurate analysis results cannot be obtained.
7. Close the cover.
8. After performing the above check, select “Yes” and press enter.
The analysis operation starts. After “STANDBY
→
MEAS. 1” is displayed on the upper
left of the screen, “MEAS. 1” is displayed.
After starting analysis, the first result is calculated. Results print as they are completed for
the sample.