Automated Chemistry Analyzer
AU2700 User's Guide
Chapter C - Basic Operations
C-81
May, 2002
To repeat a sample from a routine rack on the STAT table:
1. From the main menu, select [Parameters], [System], [System].
2. Select function key F4 (set).
3. Use the pull-down menu by the “Auto/Standard Repeat” field and select “Standard.”
4. Press function key F2 (exit) twice.
5. From the main menu select [Parameters], [Repeat Parameters], [Repeat Common].
6. Select function key F4 (set). The box to the left of the Auto Repeat field should have a
check mark in it.
7. In the [Repeat Common] screen, select any flags to generate repeat requisitions. Flags
can be set in one of the 5 following categories.
A. The normal repeat mode uses the original sample volume to repeat the test.
B. The dilution repeat mode uses a user defined sample volume smaller than the original
sample volume to perform a “dilution.” Dilution can also be performed by using a
dilution cuvette and defining the amount of dilution from 5-100. See How to Perform
a Manual Repeat Run’ for more information.
C. The “OR mode” requisitions a repeat if any flag selected is generated.
D. The “AND mode” requisitions a repeat only if all flags selected are generated on a
test.
E. The “Condense” category increases the sample volume for a repeat run.
8. If you want to set up specific test repeat ranges: Select [Parameters], [Repeat
Parameters], [Repeat Specific]. Enter a range at the “Repeat Decision Level” to generate
J (above the range) and K (below the range) flags.
9. If you want to perform a dilution refer to ‘How to Perform a Manual Repeat Run’.
10. Select [Routine], [Test Requisition], [Repeat]. Select the tests that require repeat analysis
based on flags generated from the data log printout. Select function key F6 (STAT
repeat). The sample will be assigned a repeat sample number starting with HP001,
HUP001, or HXP001.
11. Print a repeat run worklist.
12. Place the sample in a position on the STAT table designated for serum or urine repeats.
If the sample containers are barcoded, they can be placed in any order in the designated
areas of the STAT table after being requisitioned for STAT table analysis. Press the STAT
button.
See Also
For detailed information on using the STAT table, refer to the section called “Using the STAT Table:
Normal Analysis Mode.”