C-50
Chapter C - Basic Operations
Automated Chemistry Analyzer
AU2700 User's Guide
May, 2002
Setting Sample Containers into the Rack
Place the sample containers into the rack so that they meet the conditions shown in the figure
below:
Be sure to use the correct rack color.
•
When using sample containers with a barcode label, position the barcode surface within
•
+2
mm to the center of the side slit of each hole on the rack.
Instruction
Do not rotate the sample containers set in the rack to change their direction. The barcode may be
contaminated or scratched resulting in a barcode read error. Remove the sample container from the
rack, and then rotate the container in the correct direction before reinserting it into the hole.
OLU2005E
Rack
Sample cup
The rack ID label is
applied to this surface.
Direction that rack moves
The rack ID label is
applied to this surface.
Barcode label
The barcode label is
applied to this surface.
Flex Rack Adapters
The flex rack does not need adapters for sample containers with dimensions of 13.60 to
16.00 mm. This rack does need adapters for sample containers with dimensions of 11.5
to 13.6 mm. When using containers with a diameter of 14.0 to 15.0 mm, the operator will
need to determine if an adapter is necessary since plastic sample containers may be tapered
and the diameter may be difficult to judge.