C-44
Chapter C - Basic Operations
Automated Chemistry Analyzer
AU2700 User's Guide
May, 2002
Warning
In sequential analysis, set sample containers on the white rack in sample number order without
leaving any empty positions. If any position is different between the sample numbers set by the
requisition operation and sample positions on the white rack, concordance errors will occur. The
following example shows tests that were requisitioned for 15 samples:
4114
Sample Nos. determined automatically
: Sample loaded
10
1 2 3 4 5 6 7 8 9
11 12 13 14 15
Sample Nos. set by requisition operation
• If the sample cups are set without any unloaded position:
10
1 2 3 4 5 6 7 8 9
11 12 13 14 15
Tests selected by requisition operation
J
A B C D E F G H
I
K L M N O
: Sample unloaded
White rack
Sample Nos. determined automatically
10
1 2 3 4 5 6 7 8 9
11 12 13 14 15 16
Sample Nos. set by requisition operation
• If the sample cups are set with one position unload:
Sample Nos. shifted
A sample unload alarm takes place.
Analysis test F for sample No.6 is not executed.
A no-test requisition alarm takes place.
An analysis test is generated automatically.
9
1 2 3 4 5
6 7 8
10 11 12 13 14
Tests selected by requisition operation
J
A B C D E
G H
I
K L M N O
15
White rack
If the analysis mode between racks changes, the sample numbers determined automatically
according to a setting position on the rack will differ.
4115
Sample Nos. determined automatically
: Sample loaded
1 2 3 4 5
6 7 8
: Sample unloaded
White rack
Moves to Analysis 2 mode
Sample Nos. determined automatically
10
1 2 3 4 5 6 7 8 9
11 12 13
White rack
No change in mode