C-58
Chapter C - Basic Operations
Automated Chemistry Analyzer
AU2700 User's Guide
May, 2002
Tips
Condensation may appear in the reagent refrigerator under high-temperature and high-humidity
environments. Remove the condensation by wiping the reagent refrigerator with a dry cloth.
3. Set the reagent bottles in the compartment by gently pressing down on the front and back
of each bottle. Replace old bottles with new ones if the reagent level is low, or if the
reagent stability or expiration date has expired.
Warning
It is imperative that the AU2700 reagent bottles are placed correctly in the refrigerator. If the end of a
bottle is up, reagent probe problems could occur.
Reagent bottles without barcodes: Place them in the reagent positions that have been set in
•
[Reagent Status].
Reagent bottles with barcodes:
•
Once the Reagent ID code is set in [Parameters], the
reagents can be placed in any position. Bottles with barcodes cannot be placed in “fixed”
positions set in [Reagent Status].
Warning
When setting 15 mL or 30 mL reagent bottles, use a partition to secure the bottle. If using 120 mL or
180mL bottles, remove the partition between bottles. Two 60 mL positions are required for the 120 mL
bottles and three 60mL positions are required for the 180mL bottles.
4 .Replace the R1 or R2 reagent refrigerator cover to its original position.
When installing
the reagent refrigerator cover, make sure that the positioning pin on the reagent
refrigerator is inserted in the hole in the reagent refrigerator cover. If the cover is not
seated properly, a “Reagent Cover Open” alarm occurs.
5. Select the [System Status] icon.
The [System Status] screen is displayed.
6 .Select [Reagent Status].
A screen for checking the reagent status is displayed. When this
screen is displayed, the reagent check status is “Unchecked.”
7 .Press function key F5 (check start).
The reagent check start window is displayed.
8. Select “Check for all positions.”
The amount of remaining reagent is automatically
checked. After the reagent has been checked, the reagent check status is displayed as
“CHECKED.”