D-172
Chapter D - Software
Automated Chemistry Analyzer
AU2700 User's Guide
May, 2002
4.8 Standby Set
(A)Auxiliary
(S) Standby Set
ALARMCLEAR
?
WARMUP
Help
Exit
Standby Set
11/30/2000 10:10
OLP1stand
Switch to Standby mode?
Because Warm Up is bypassed,
the system may not be thermally equilibrated.
Yes No
Description:
Use this screen to shift the analyzer to Standby from a warm-up mode. If the system is in a stop
mode, shift to standby using the stop key or icon (square located at the top of the screen). If
an emergency stop is performed, the analyzer stays in a warm-up mode for 1.5 hours. Use this
screen to bypass the warm-up time if necessary.
Caution
If analysis is performed after shifting to the Standby Mode, results could be compromised since the
system temperature did not have time to stabilize completely.
Use this software screen to:
Shift to the Standby Mode:
For procedures, refer to "Standby Mode" located in the Basic Operations Chapter.