Automated Chemistry Analyzer
AU2700 User's Guide
Chapter C - Basic Operations
C-9
May, 2002
(P) Parameters
(A) Auxiliary
(M)Maintenance
(L) Logout
(R) Routine
(U) User
ALARMCLEAR
?
1997/09/03 10:10
WARM UP
OLU1010E
Login Name
Cancel
Password
OK
Login
Note
The system security pop-up screen appears only if the analyzer options are activated. Enter the log-in
name and password then press enter.
See Also
For information about creating passwords, refer to the Software Chapter.
6. The window closes and the main screen appears.
See the illustration below for the
location of the menu bar.
(P) Parameters
(A) Auxiliary
(M)Maintenance
(L) Logout
(R) Routine
(U) User
ALARMCLEAR
?
WARM UP
1997/09/03 10:10
OLP1042
Menu bar
7. Perform the following daily maintenance procedures. Refer to the Maintenance Chapter
for detailed procedures.
Check the sample and reagent syringes for leaks or condensation.
•
Fill the concentrated wash solution bottles.
•
Analysis pauses and a warning message
displays on the screen if there is insufficient concentrated wash solution.
Check the sample probes, reagent probes and mix bars.
•
The probes should be straight
and clean. The mix bars should be clean with no chips in the Teflon. Select the following
menus: [Maintenance], [Analyzer Maintenance], H. Prime Washing-line. Press the
STAT/Rotation Diag. button. Verify that a thin, straight stream of diluent dispenses from
all probes. If the probes or mix bars appear dirty, clean them with an alcohol prep.
Verify that the printer “Ready” light is on and that there is sufficient printer paper.
•
Change DI water in the pre-dilution bottle.
•