D-170
Chapter D - Software
Automated Chemistry Analyzer
AU2700 User's Guide
May, 2002
4.7 Calibration Verification
(A)Auxiliary
(V) Calibration Verification
1.11/30/2000 12:30
ALARMCLEAR
?
STANDBY
Help
Exit
Test No.
Calibration Verification
11/30/2000 10:10
Index:
Test Name.: 1. Alb
Material No.: 1
Sample No.
Sample ID
Display
Chart
Set
S. ID
OLP1calv
Level 1:
Level 2:
Level 3:
Level 4:
Level 5:
Level 6:
Type:
Serum
Select Sample Material Prm.
Description:
Use this screen to select the data to be used in generating a calibration verification graph.
Samples must be processed on the analyzer first.
Index:
Use the pull-down menu to select a different date and time.
Test Name:
Use the pull-down menu to select the test name.
Type:
Select the sample type by using the pull-down menu.
Material No.:
This field defaults to 1. To change the material number, use the keyboard or </>
buttons.
Sample No.:
Set the sample number range within the numbers printed in parenthesis. To
specify all samples, enter an asterisk (*). To leave the field blank, enter a space.
Sample ID:
Set the sample ID range within the range specified in [Parameters], [Requisition
Format]. To specify all samples IDs, enter an asterisk (*). To leave the field blank, enter a
space.