Automated Chemistry Analyzer
AU2700 User's Guide
Chapter D - Software
D-143
June 16, 2006
3.9 Special
3.9.1 Contamination Parameters
F4
OLP3046
F3
Referring to a
contamination
table
Setting a
contamination
table
Outputting the
setting contents in
a data list
(P)Parameters
(C)Special
(C)Contamination Parameters
One of the tabs
can be selected
Button function
Described in
another section.
Press this key.
Main operation flow
Operation flow
ALARMCLEAR
?
STANDBY
Help
Exit
Set
Contamination Parameters
11/30/2000 10:10
OLP1cpa
No. Preceding Following Reagent Probe Wash Cancel Mixer Cuvette
Test Name Test Name Cleaner Kind Count
1.
2.
3.
4.
5.
6.
7.
8.
9.
10.
11.
12.
13.
14.
Contamination Carry Over - 1 Carry Over - 2
Periodic Cleaning
F4 (Set)
F5 (Edit)
Set Contamination
Close
No.:
Preceding Test Name:
Following Test Name:
Cleaner Kind:
Wash Count:
Cancel
Mixer
Cuvette
ALARM CLEAR
Help
Exit
Edit
Description
Use this screen to program a decontamination protocol when reagent cross-contamination is
suspected. Enter the suspected test(s) in the “Preceding Test Name” and the affected test in the
“Following Test Name”. Choose what action the instrument should take if the identified tests
are run consecutively.
Use this software screen to:
Set a Contamination Table
A maximum of 100 test combinations can be defined in contamination parameters.
1. Select [Parameters], [Special], [Contamination Parameters].