D-60
Chapter D - Software
Automated Chemistry Analyzer
AU2700 User's Guide
February 4, 2004
2.4.3 Reagent Blank Monitor
OLP3006
F3
F6
F3
F5
F8
F5
F7
F3
F8
Selecting an
analysis test
Referring to the
reagent blank data
(R) Routine
(C)Calibration Monitor
(B)Reagent Blank Monitor
Outputting the
reagent blank data
to a data list
Referring to a
registered comment
Specifying another
analysis test
Referring to a record
on a graph
Setting the display
conditions
Referring to the
record data
Registering a
comment to an
analysis test
F6
F4
One of the tabs
can be selected
Button function
Described in
another section.
Press this key.
Main operation flow
Operation flow
"Reagent Blank Monitor" (Select) screen
"Reagent Blank Monitor" (Display data) screen
ALARMCLEAR
?
11/30/2000 10:10
STANDBY
OLP1rbm
Reagent Blank Monitor
Type:
Serum
Change
Test
Help
Exit
Alb ALP ALT AST CO2 Tbili
Tbilb Dbili Dbilb Cal Chol HDL-C
CK CK-MB Creat GGT Gluc Phos
Iron LDH LIP Mag Tpro Trig
UIBC BUN Uric S-CAL S-CRE S-GLU
Upper: Inner
Lower: Outer
Description:
Use this screen to review the current and past reagent blanks during reagent troubleshooting.
These records can be reviewed either on a graph or from the data. Also, the data can be printed
in a list. ISE tests and LIH tests cannot be monitored here.
Type: The sample type can be changed to serum, urine, or other. Select the sample type from
the pull-down menu.
Test - List: Use the mouse to deselect and select tests.
Use this software screen to:
Display reagent blank data:
For each analysis test it is possible to refer to the reagent blank data for every absorbance read
point (0 to 27). If R1 is less than 150, read point 11 to 27 will display. Select an
analysis test. The following operations are permitted:
Display reagent blank data for a different test
Use the pull-down menu located by the test name field.
Display reagent blank data for a different date
Use the pull-down menu located by the "RB Measured Date" field.
Display a history of reagent blank data:
Press function key F6 (history) to graphically display the last 30 reagent blanks.