D-106
Chapter D - Software
Automated Chemistry Analyzer
AU2700 User's Guide
May, 2002
3.4.2 Repeat Specific
Referring to
individual repeat-run
parameters
OLP3059
F4
F3
F8
F8
Setting individual
repeat-run
parameters
(P) Parameters
(R)Repeat Parameters
(I)Repeat Specific
Outputting the
setting contents in
a data list
Specifying another
analysis test
One of the tabs
can be selected
Button function
Described in
another section.
Press this key.
Main operation flow
Operation flow
ALARMCLEAR
?
STANDBY
Help
Exit
Test No.
Repeat Specific
11/30/2000
10:10
Test Name.: 1. Alb < >
Sample Volume: 1.0 ul
Diluted
Pre-Dilution Rate: 1
Condense
Normal
OLP1rspec
Type:
Repeat Decision Level L: -999999.9
H: 999999.9
Dilution Volume: 0 ul
Sample Volume: 1.0 ul
Pre-Dilution Rate: 1
Dilution Volume: 0 ul
Sample Volume: 2.0 ul
Pre-Dilution Rate: 1
Dilution Volume: 0 ul
Set
Serum
Description:
Use this menu to set the diluted repeat sample volume, the repeat range to generate repeat
requisitions, and automatic repeat overwrite for each test. J and K flags will be generated if the
repeat range is exceeded in this screen.
Diluted, Normal, and Condensed Volume, and Rates:
Use one of the following options for
first time analysis:
First time
analysis
Undiluted analysis
Pre-diluted analysis (dilution rate: 1, 5 to 100 times)
OLU4112E
Use one of the following options for repeat analysis:
Repeat-run
analysis
Same vol. sample
analysis
Decreased sample
analysis
Decrease the quantity of analysis
sample that has the same dilution
rate (1, 5 to 100 times) as that of
the first time analysis sample.
Increased sample
analysis
Increase the sample dilution rate
more than that of the first time
analysis sample.
Increase the quantity of analysis
sample that has the same dilution
rate (1, 5 to 100 times) as that of
the first time analysis sample.
Decrease the sample dilution rate
less than that of the first time
analysis sample.
OLP4113
: Analysis using smaller amount
of sample than that for the first
time analysis.
: Analysis under the same condition of sample
quantity as that of the first time analysis.
: Analysis using greater amount
of sample than that for the first
time analysis.