Automated Chemistry Analyzer
AU2700 User's Guide
Chapter D - Software
D-51
May, 2002
Search for a sample number (QC, Cal and/or Reagent Blank): Enter a sample number range.
Note for QC Samples: Numbers for serum, urine and other are user defined.
Note for Calibration Samples: Numbers for serum, urine and other are user defined.
For calibration and reagent blank samples, select a cuvette count in a range between 1 and 4,
then press the Enter key. The cuvette count number is specified in [Parameters], [Calibration],
[Calibration Specific]. Specify which data is to be used from the photometry results obtained
from 4 cuvettes. The selected value will be highlighted on the screen.
5. Press function key F5 (Search Data).
Information about the reaction process searched
according to the specified condition will be displayed as a graph.
Reaction Monitor” (Display curve) screen
After a message to show the search is in progress is displayed, the reaction process will be
displayed as a graph. If no corresponding data exists, a message to show the fact will be
displayed. Dotted lines for representing the main wavelength, sub-wavelength and reaction data
on the graph can be identified by each display color.
ALARMCLEAR
?
STANDBY
Reaction Monitor
1997/09/03 10:10
Index:
Sample No.:
nnnnnnn
Wave length 1: nnn
Wave length 2: nnn
Data
nnnnnnn
nnnnnnn
nnnnnnn
nnnnnnn
nnnnnnn
0
5 10 15 20 25 27
POINT
OD
Help
Exit
Display
Data
Rgt. Inf.
RB/Cal.
Axis
Change
QC, Cal/Count:
Cuvette:
Meas.Type:
Sample ID:
Test:
Result:
Cuv./Mix. No.:
Reagent No.R1:
R2:
Method:
Point:
1:
2:
Measured Time:
Reagent Blank:
Photocal:
OLP1211
Cuvette:
Displays either “Inner” or “Outer”.
Sample No. - left box:
The applicable sample No. will be displayed.
Sample No. - right box:
the case of a repeat-run sample, S.No. of the original sample will be
displayed.
QC, Cal./Count - left box:
The control No. will be displayed for a control sample, and the
calibrator No. will be displayed for a calibration sample. The sample type will be also displayed.
Example: “3 (Serum)” or “5 (Urine)”:
the control No. or calibrator No. shows the samples
are all serums, the sample type will not be displayed.
QC, Cal./Count - right box:
the cases of calibration samples or reagent blank samples, the
cuvette count will be displayed in this box.