Automated Chemistry Analyzer
AU2700 User's Guide
Chapter C - Basic Operations
C-35
June 16, 2006
2.5.6 Real-time Downloading of Requisitions from a Host Computer
To automatically download requisitions from a host computer, the system must be set to “Real-
time” in [Parameters] [Online]. The T.R.I. Receive prompt can be set to Real-time, Batch,
or none. Set to “real-time” to automatically download requisitions according to the following
procedure:
1. From the main menu, select [Parameters], [Online] in that order.
2. Select F4 (set).
3. Select the pull-down menu in each category on screen to select “Real-time” at any of the
“Test Requisition Information” fields for automatic downloading by sequence or sample
ID.
Tip
If an online error occurs during real-time online analysis, the analyzer defaults to test profile number 0.
Note
Manual requisitions have a higher priority than online requisitions. The barcode identification
can be entered in the sample ID field, then select the desired tests following test requisition. The
system will automatically process the barcoded sample with the revised requisition.