D-190
Chapter D - Software
Automated Chemistry Analyzer
AU2700 User's Guide
May, 2002
Rack Buffer
N/Reset:
Resets the rack buffer belt.
O/Forward Step:
Steps the rack buffer belt forward 1 step.
P/Back Step:
Steps the rack buffer belt back 1 step.
Select the tab located in the upper left corner of the screen called Sample Set Unit 2.
Rack Collection 1
A/Reset:
Resets the rack collection area.
B/Step 1 (1 Pitch):
Moves the rack collection push bar toward the rack collection area and
then back when the Rev./Diag button is pressed.
Rack Collection 2
C/Reset:
Resets the repeat run bar.
D/Step:
Moves the repeat run bar behind the rack collection area.
Rack Receiver
F/Reset:
Resets the rack receiver area.
G/Step:
Steps the rack receiver push bar.
Rack Carrier
H/Reset:
Resets the selected rack bar.
I/Forward Step:
Steps the rack bar in the selected lane.
J/Collect Step:
The rack bar, in the selected lane, moves forward and then backward when
the Rev./Diag button is pressed again.
K/Belt Rotation:
Moves the belt backward and then forward when the Rev./Diag button is
pressed again.
Select the tab located in the upper left corner of the screen called Combination.
Photometer Check
A/0-100%:
Performs a 0% and 100% photometer check on a specified cuvette. The check
is performed at 13 wavelengths.
B/0-100% Sequence:
The specified cuvette is measured 20 times. Results are calculated
from the average value, range, SD & CV. Results are printed.
C/Rotation:
Performs a 100% Photometer Check on cuvettes 1 to 165 on both the inner and
outer wheels, using all wavelengths as primary and secondary.
D/Rotation Sequence:
Performs a 100% Photometer Check on cuvettes 1 to 165, using all
wavelengths as primary and secondary, measured 20 times.
Whole
E/Reset:
Aligns all mechanical parts to their home position.
F/Magnet Off:
Removes all power from pulse motors. To reset select "E/Reset."
Barcode Read
G/S. ID:
Reads the sample ID at the cup position.
H/S. ID Movement:
Performs a complete rack sequence and reads S. ID data.
I/Rack ID:
The rack located in the "rack ID read position" is read.
J/Rack ID Movement:
The barcode reader is activated.