D-54
Chapter D - Software
Automated Chemistry Analyzer
AU2700 User's Guide
May, 2002
B. Enter a test number of an analysis test to be searched and select “OK.”
After searching data, the following appears:
F7: Change Axis (Set display conditions):
By setting the conditions to display the Y-axis of a graph, it is possible to modify the display
size of a reaction process curve on the graph.
1. After selecting [Routine], [Reaction Monitor], press function key F5 (search data) and the
[Reaction Curve] screen appears. In the [Reaction Curve] screen, press function key F7
(change axis).
A window to set-up the display conditions is displayed.
2. Set the display scale to either “Manual” or “Auto.” If “Manual” is selected, set the lower
and upper limit values.
A specific part of the graph will be magnified and displayed.
If
“Auto” is selected, the largest possible graph will be displayed within an area in which all
data can be contained.
Graph Display: The X axis displays the measuring point. The Y axis displays the
measured OD value.