D-66
Chapter D - Software
Automated Chemistry Analyzer
AU2700 User's Guide
May, 2002
Test Name, No., QC Monitor: Displays the test name, inner, and/or outer cuvette wheel and
position and QC monitor information. The QC monitor field displays the following:
V
oid
:
n
o
data
*:
1 SD
1:
O 1 SD and 2 SD
2:
O 2 SD and 3 SD
3:
O 3 SD
E:
Error, no data value
Use this software screen to:
Check the daily variation chart:
For procedures, refer to "Checking Quality Control" located in the Basic Operations Chapter.
Display a specific control sample:
1. Select [Routine], [QC Monitor], [Daily Control].
2. Select a sample type using the pull-down menu located by the "Type" field.
3. Select function key F8 (control no.), enter a control number of the control sample to view.
Select "OK."
4. Select a test, then press function key F6 (daily data).
A daily variation chart related to
the specified control sample will be displayed.
5. To display the sample number, measuring type, control ID, result, flag, and lot number,
press function key F7 (daily data). Select "Close" to exit the window.
6. Press function key F2 (exit) twice.
Add comments for the displayed daily variation data:
A comment can be added to each control sample.
1. Press function key F4 (comment).
A window to register a comment is displayed.
2. Enter a comment. A maximum of 50 characters can be entered. To choose from a list of
comments, select the comment button and highlight a comment with your mouse.
3. Select “OK.”
The entered comment will be saved. If “Cancel“ is selected, the comment
will not be saved and the window will be closed.
Set the display size:
To modify the display size of a graph, change the X axis.
1. Select [Routine], [QC Monitor], [Daily Control].
2. Select the tests displayed on screen using your mouse.
3. Press function key F6 (daily data).
4. Press function key F8 (change axis).
A window to set the display condition appears.