C-12
Chapter C - Basic Operations
Automated Chemistry Analyzer
AU2700 User's Guide
May, 2002
2.2 Starting the ISE
Perform the following ISE start-up and daily maintenance procedures before beginning analysis.
For information on commercial control sera for the ISE, refer to Section 1 at the beginning of
this chapter.
Starting the ISE Unit
1. Verify the ISE power switch is on.
2. Verify that the ISE reagents have not expired. Refer to the package insert for information
on open bottle stability.
OLU2002E
FRONT
ISE POWER
ON
OFF
ISE power switch
ISE reagent dispenser
REF solution tank
Buffer
solution
tank
MID solution tank
Check the level of liquid in the tanks. If the liquid level is low, an insufficient liquid alarm will
sound. After this alarm occurs, the following number of samples can be processed: 180 for
Mid, 600 for Reference, and 240 for Buffer solution.
See Also
For more information about changing reagents, refer to the Maintenance Chapter.
3. Verify the ISE buffer syringe is not leaking and that condensation is not visible.
ISE
OFF
OLU2065
FRONT
ISE power switch
Fixing screws
Case head
Syringe case
: Possible leakage locations