Automated Chemistry Analyzer
AU2700 User's Guide
Chapter D - Software
D-19
May, 2002
1.5 Analyzer (ANL) Status
Referring to the data
for samples under
analysis
F5
Referring to main
information on
system status
[System Status]
(4) Data
Display
Referring to the data
for all samples
Referring to the data
for ISE (option)
Referring to the data
for samples analyzed
with the STAT table
Select [All Data] tab
"Data Display" screen
Select [STAT Data] tab
Select [ISE Data] tab
F6
F5
Referring to data
related to analyzed
smples
F5
OLU3013E
One of the tabs
can be selected
Button function
Described in
another section.
Press this key.
Main operation flow
Operation flow
ALARMCLEAR
?
STANDBY
Help
Unit-1
Sampler
RACK
START ---
D. I. WATER
DET. CONC. B
DET. B
ISE
NOT USE
Auto Rerun
Unconnect
Electrode
MID.
BUFFER
REF.
Exit
ANL Status
1997/09/03 10:10
OLU1107E
Unconnect
Description:
This screen displays the status of the following:
Rack feeder status
Detergent (Wash Solution)
ISE reagent status
DI water tank level
R1/R2 refrigerator temperature
Incubation Bath Temperature
Cuvette wheel status
Reagent Compartments
Vacuum Tank
It also displays the sample number of the sample being analyzed in the R1, sample and R2
position.
See Also
For information on correcting many of the following abnormal messages, refer to the “Fatal and Nonfatal
Errors” section located in the Flags Chapter.
Use this software screen to:
Check Analyzer Status:
Move the cursor onto the [ANL status] column, then press the Enter key.
The [ANL Status]
screen is displayed.
Operation mode:
m
essaGe
m
eaninG
c
hanGe
in
bacKGround
color
AUTO (Ready)
Analysis is possible.
Blue
STOP
Analysis is not possible.
Red or Yellow
STANDBY
Analyzer is standing by.
Blue
MEASURE
Analysis in progress.
Red
W1
W1 in progress.
Red
Содержание AU2700
Страница 17: ...A 4b Automated Chemistry Analyzer AU2700 User s Guide November 1 2009 This page intentionally left blank ...
Страница 178: ...D 4 Chapter D Software Automated Chemistry Analyzer AU2700 User s Guide March 31 2005 ...
Страница 260: ...D 86 Chapter D Software Automated Chemistry Analyzer AU2700 User s Guide May 2002 ...