Automated Chemistry Analyzer
AU2700 User's Guide
Chapter D - Software
D-69
May, 2002
3. Highlight the tests to be displayed or printed.
4. Select function key F6 (day to day) to view the graphs and statistics of the indexes
selected. Each index will be an average QC value.
5. To print, select function key F3 (print). Highlight the printing option you want to perform:
day to day, daily data, graph, graph plus comment, or all data plus comment.
6. Select “OK” to start printing.
7. Select function key F2 (exit) twice.
Display daily statistical data:
1. Select [Routine], [QC Monitor], [Day to Day Control].
2. Set the index using the pull-down menu next to the "index" field.
3. Highlight the tests to be displayed or printed.
4. Select function key F6 (day to day) to view the graphs and statistics of the indexes
selected. Each index will be an average QC value.
5. Press function key F7 (daily data) to refer to statistical values, such as the number of
samples, mean, and standard deviation of the daily data. Items for which daily statistical
values are displayed include the sample number and analysis result.
F8: Control No.:
Use this screen to change the control number currently viewed.
1. Press function key F8 (control no.).
2. Enter the control number to be viewed. A number from 1-60 can be entered. The default
analyzer settings are: 1-30 for serum, 31-45 for urine, and 46-60 for others. These
settings are user defined and can be changed by authorized personnel.
3. Select "OK."