Automated Chemistry Analyzer
AU2700 User's Guide
Chapter D - Software
D-129
May, 2002
Use this software screen to:
Set STAT Calibration Positions:
1. Select [Parameters], [Calibration], [STAT Table Calibration].
2. Select the round name using the pull-down menu.
3. Press function key F4 (set). The cursor appears in the "Operation" field. Select "Auto,"
"Manual," or "None."
4. Press function key F5 (edit).
The "Calibrator setup" dialog box is displayed.
5. Use the arrows at "Calibrator Name" to program a calibrator material to a position on
the STAT table. The claibrator material must have a previously assigned Calibrator No.
programmed in [Parameters], [Calibration], [Calibrator].
6. To change the position number use the "<" or ">" buttons located on screen. Repeat step
5 for each position that needs to be set.
7. After completing the settings, press function key F2 (Exit).
The settings are saved.
Set STAT Table Parameters:
1. Select [Parameters], [Calibration], [STAT Table Calibration].
2. Select the "Test" tab located in the upper left corner of the screen using your mouse.
3. Select the round name using the pull-down menu.
4. Press function key F4 (set).
5. Select the test(s).
6. Press function key F5 (edit).
7. Set the following:
Available/Unavailable:
Displays whether a test is enabled (available) or disabled (unavailable).
Calibration Mode:
Use your mouse to select the pull-down menu. Select one of the following:
All+RB, One Point+RB, or RB.
Execution Type:
The following selections are available: Change Bottle No. or Change Lot No.
Use "Change Bottle No." when the reagent bottle # of a test you have selected has changed. Use
"Change Lot No. when the reagent lot # of a test you have selected has changed.