C-80
Chapter C - Basic Operations
Automated Chemistry Analyzer
AU2700 User's Guide
May, 2002
4.3.5 How to Perform a Manual Repeat Run Using the STAT Table
To quickly perform repeat runs on a small number of samples, use the STAT table. The
following procedure explains how to program and perform a manual repeat run. If the system
has been preprogrammed to perform a manual repeat run, follow the procedures in the AU2700
Quick Response Guide.
To repeat a priority STAT sample on the STAT table:
1. After running priority STAT samples, print a repeat run worklist.
2. From the main menu, select [Parameters], [System], [System].
3. Select function key F4 (set).
4. Use the pull-down menu by the “Auto/Standard Repeat” field and select “Standard.”
5. Press function key F2 (exit) twice.
6. From the main menu select [Parameters], [Repeat Parameters], [Repeat Common].
7. Select function key F4 (set).
The box to the left of the Auto Repeat field should have a
check mark in it.
8. In the [Repeat Common] screen, select any flags to generate repeat requisitions. Flags
can be set in one of the 5 following categories.
A. The normal repeat mode uses the original sample volume to repeat the test.
B. The dilution repeat mode uses a user defined sample volume smaller than the original
sample volume to perform a “dilution.” Dilution can also be performed by
using a dilution cuvette and defining the amount of dilution from 5-100. See
‘How to Perform a Manual Repeat Run’ for more information.
C. The “OR mode” requisitions a repeat if any flag selected is generated.
D. The “AND mode” requisitions a repeat only if all flags selected are generated on a
test.
E. The “Condense” category increases the sample volume for a repeat run.
9. If you want to set up specific test repeat ranges: Select [Parameters], [Repeat
Parameters], [Repeat Specific]. Enter a range at the “Repeat Decision Level” to generate
J (above the range) and K (below the range) flags.
10. If you want to perform a dilution refer to ‘How to Perform a Manual Repeat Run’.
11. Place the STAT sample in a position on the STAT table for STAT serum repeat, STAT
urine repeat, or STAT other repeat. If the sample containers are barcoded, they can be
placed in any order in the designated areas of the STAT table. Press the STAT button to
begin analysis. The first repeat sample number will be HP001, HUP001, or HXP001.