D-78
Chapter D - Software
Automated Chemistry Analyzer
AU2700 User's Guide
May, 2002
Sample ID:
Enter the sample ID range according to the range set in [Parameters], [Requisition
Format]. Enter an asterisk (*) to designate all sample IDs.
F5 : Correct Data
Analysis results can be corrected according to a coefficient included in the calibration formula. A
“c” is appended to the corrected value, indicating an abnormal data symbol.
1. Specify a search key, then press the F5 (Correct Data) key. The “Set Correction Factor”
dialog box is displayed.
2. Enter coefficients A and B of the following calculation formula using each number
between -9999999 and +9999999. Y (data after correction) = AX (data before correction)
+
3. Select the “YES” button. A message is displayed informing user Data correction has
started. After the correction has been completed, a completion message appears. Confirm
the message, then select the “OK” button.
To interrupt the correction, click the “Cancel” button during correction.
F6 : Recal. Data
Concentration values can be re-calculated based on the stored absorbance values. As a result,
the existing analysis data will be overwritten by the data after re-calculation.
1. Press the F6 (Recal. Data) key. The “Re-calculation” dialog box is displayed.
2. Select the “YES” button. A message is displayed informing user Data Re-calculation has
started.
To interrupt re-calculation, click the “Cancel” button during re-calculation.
Tip
Data saved as an OD value can be modified by performing re-calculation.
Data saved as concentration values can be corrected by performing recalculation. Correction is not
possible for data in which no OD values have been saved.
F7 : Search Add
In addition to a sample type and analysis test, patient information, such as sex, age, etc., can be
specified as a search key for data search.
1. Press the F7 (Search Add) key. The “Search Add” dialog box is displayed.
2. Set “Sex”, “Age’s upper limit”, “Age’s lower limit” , “Cuvette” and 6 types of “Patient
information”.
3. After the entry has been completed, select the “YES” button. The setting contents will
be registered as the search keys.