2.4.4 Performing Auto-calibration and QC on the STAT Table
Important
Laboratories that use the auto-calibration feature must be sure that
all
on-board test parameters are
programmed for auto-calibration and that
all
calibrators are present in their programmed positions on the
STAT carousel during operation. Failure to ensure adequate amount and quality of calibrator materials
could cause inaccurate results.
Reagent blanks, calibrations, and QC are performed automatically based on programming from
[Parameters].
Caution
If STAT table QC is set to “Auto” and sample blank tests are requisitioned, set the cyclic type for the
color and blank test to “Sample.” If it is set to “Test,” optimal results will not be obtained.
Note
[Stat Table Calibrations] and [STAT Table QC] in [Parameters] must be programmed to “Auto.”
1. From the main menu select the [System Status], [STAT Table Status].
2. Load the reagent blank, calibrators, or QC cup(s) in the correct positions on the STAT
table. Use the STAT ROTATION/DIAG Button if necessary.
Only place calibrator and
QC material on the STAT table if analysis from this material is required.
3. Perform a STAT table check:
A. From [STAT Table Status], place reagent blank, calibrator, and QC material in the
correct positions on the table. Use the STAT ROTATION/DIAG Button to turn the
table if necessary.
B. Select function key F5 (check start). In the STAT Check Window, select “Yes” at the
“Start Checking?” prompt.
Note
The table rotates and senses if reagent blank, calibrator, and/or QC cups are loaded. Alarms are
generated if the positions are defined for reagent blank, calibrator, or QC in [System Maintenance] and
cups are not detected.
4. Reagent blank, calibration and QC are performed automatically based on programming
from the [Parameters] menu.