Barcode check character
b
arcode
c
hecK
character
,
method
NW7
Least significant digit, MODULA-16
CODE39
Least significant digit, MODULA-43
CODE128
Least significant digit, MODULA-103
2 of 5 interleaved
Least significant digit, MODULA-10
Number of barcode digits
•
Maximum 26 digits (all)
ID Search Type
•
Real-time on-line entry
Batch Online
Width of bar and space
•
NB (narrow bar) is standard. If you require a different bar width and space refer to the
following specifications:
nb (n
arrow
b
ar
)
ns (n
arrow
s
pace
)
wb (w
ide
b
ar
)
ws (w
ide
s
pace
)
G (G
ap
)
0.2 to 0.5 mm
NB to 1.25NB
2.0NB to 3.0NB
2.0NB to 3.0NB
NB to 3.0NB
0.165 < 0.2 mm
NB to 1.25NB
2.2NB to 3.0NB
2.2NB to 3.0NB
NB to 3.0NB
If a high print contrast signal value is obtained by printing a barcode clearly on a label with a
high reflection factor and if the width of bars and spaces are retained as follows, the narrow
barcode can be used only for NW7.
nb
ns
wb
ws
G
0.130 - 0.156mm
0.216 - 0.238mm
0.401 - 0.413mm
0.465 - 0.492
0.465 - 0.492mm
Caution
It is recommended to use bars and spaces of 2.5 NB or more so that WB (wide bar), WS (side space),
and G (gap) can be identified easily. If less than 0.2 mm of NB (narrow bar) is used, it is recommended
to use the barcode check character to retain the reliability of the barcode read.