Automated Chemistry Analyzer
AU2700 User's Guide
Chapter C - Basic Operations
C-111
May, 2002
6.8 Changing Calibrator Set Points
1. Select [Parameters], [Calibration], [Calibration Specific].
2. Use function key F8 (test no.) to get to a specific test number, or use the arrow buttons
next to the test name field.
3. Select function key F4 (set). Enter the new calibrator concentration at the “CONC” field
from the package insert.
4. Select function key F2 (exit) to set the new concentration.