D-18
Chapter D - Software
Automated Chemistry Analyzer
AU2700 User's Guide
May, 2002
Prime the ISE:
The operator can drain or add liquid from/to the ISE tubing. The following procedures can be
performed in this state:
Electrode replacement
•
P tube exchange prime
•
All (total) prime
•
Buffer prime
•
MID standard/Reference prime
•
Prime (bypass)
•
1. Select [System Status], [ISE Status].
2. From the “Result Cal,” “Slope Chart” or “Selectivity Check Results” tab, press function
key F6 (prime).
A window to select a prime operation item is displayed.
3. Using your mouse, select from the following primes to be performed:
A/Replace Electrode
•
B/P Tube Exchange Prime
•
C/Total (All) Prime
•
D/Buffer Prime
•
E/Mid/Ref Prime
•
F/Prime Bypass
•
4. Press the STAT ROTATION/DIAG Button.
The prime operation starts.