C-88
Chapter C - Basic Operations
Automated Chemistry Analyzer
AU2700 User's Guide
May, 2002
4.7 How to Reprint Reagent Blank, Calibrator, QC and
Patient Data
Reprinting Data
The analyzer is in Standby Mode:
1. On the main screen select [Routine], [Data Report], and [Data List], in that order.
The
screen for selecting the sample whose data is to be printed is displayed.
2. Select the “Patient” tab or “QC/Cal” tab located at the top of the screen.
3. Specify the index and type of the sample to be printed.
4. Enter the range of sample numbers.
It is also possible to enter the range of sample IDs.
5. Press function key F3 (print).
Start Printing?
Select “OK.”
6. After the data is printed, select function key F2 (exit).
The analyzer is in Measure Mode:
If a printer error occurs during a measure mode: The data that did not print will be stored in
the printer buffer. The mode will display as Meas. to Meas. End. (The analyzer is not in the
Standby mode.)
1. Correct the printer error.
2. Select [System Status], [DPR Status].
3. Select function key F4 (printer control).
This only displays when a problem occurs and
there is data to print.
4. A window appears to select “Resume Printing” or “Cancel Print.”
5. All data will print if “Resume Printing” is selected.