Automated Chemistry Analyzer
AU2700 User's Guide
Chapter D - Software
D-93
Automated Chemistry Analyzer
AU2700 User's Guide
Chapter D - Software
D-93
April 18, 2008
3.2 Specific Test Parameters
Setting the number
of digits in the
decimal place
OLP3009
F4
F2
F2
F2
F2
F4
F4
F4
F7
Referring to an
ISE (optional unit)
test
Referring to a
standard value
range
Setting a
standard value
range
F3
F8
F8
F8
F8
Specifying another
analysis
test
(P) Parameters
(I)Specific Test Parameters
Setting a
biochemical test
Setting an LIH test
Setting an
ISE (optional unit)
test
Outputting the
setting contents in a
data list
One of the tabs
can be selected
Button function
Described in
another section.
Press this key.
Main operation flow
Operation flow
Test Name:
Type:
Serum
Operation:
YES
General
LIH
ISE
Range
ALARMCLEAR
?
STANDBY
Help
Exit
Set
Test No.
Specific Test Parameters
11/30/2000
10:10
Sample: Volume
10.0 ul
100 ul
100
480
0
ul
%
Reagents: R1 Volume
R2 Volume
Wavelength
Method:
Reaction slope:
Measuring Point 1:
Measuring Point 2:
Linearity Limit:
No-Lag-Time:
Pri.
First
First
L
Dilution
10 ul
50 ul
50
340
27
ul
Dilution
Dilution
Sec.
Last
Last
Pre-dilution Rate:
100
0.0000
Min.OD
Max.OD
-2.0000
First L
Last L
Reagent OD limit:
-9999.999
L
Dynamic range:
1.000000
A
H
0.0000
2.5000
First H
Last H
9999.999
H
0.000000
B
100
Onboard Stability Period:
Correlation factor:
OLP1001
Description:
Use this screen to set specific parameters for each of the individual tests. Enter specific
parameters based on the reagent manufacturer’s methodology sheet. Beckman Coulter
methodology sheets are located in the User’s Guide Vol. II.
Use this software to:
View and/or Program Specific Test Parameters
1. Select [Parameters], [Specific Test Parameters].
2. Select a test by using the pull-down menu or the ►◄ buttons to the Test Name Field.
3. Select on of the following tabs: General, LIH, ISE, or Range.
Select the General Tab
Enter Specific Test Parameters
Select function F4 (Set) to program parameters for the selected test.
Test Name
Use the pull-down menu or </> to select a test name
Type
Use the pull-down menu to select “serum,” “Urine,” or “other
Operation:
Select "Yes" to perform the test for each sample type, or "No" not to.
Sample Volume
Sample Type
Serum or Other
2.0 to 50 uL (in 0.5 uL increments)
Urine
1.0 to 50 uL (in 0.5 uL increments)
Sample Dilution
All sample types: Enter 0 or 10 uL.
Note: If a sample dilution volume of 10 uL is entered, the maximum
sample volume is 25 uL.
R1 Volume and
Dilution
R1 Volume:
Enter 25 to 300 uL (in 1.0 uL increments).
Dilution:
Enter 0 or 10 to 250 uL.
R2 Volume and
Dilution
R2 Volume:
Enter 0 (no R2 reagent) or 25 to 300 uL (in 1.0 uL increments)
Dilution:
Enter 0 or 10 to 250 uL.
Wavelength Pri.:
Select 340, 380, 410, 450, 480, 520, 540, 570, 600, 660, 700, 750, 800. Each
assay must have a primary wavelength.
Содержание AU2700
Страница 17: ...A 4b Automated Chemistry Analyzer AU2700 User s Guide November 1 2009 This page intentionally left blank ...
Страница 178: ...D 4 Chapter D Software Automated Chemistry Analyzer AU2700 User s Guide March 31 2005 ...
Страница 260: ...D 86 Chapter D Software Automated Chemistry Analyzer AU2700 User s Guide May 2002 ...