Automated Chemistry Analyzer
AU2700 User's Guide
Chapter D - Software
D-45
February 4, 2004
2.2.4 Calibration
One of the tabs
can be selected
Button function
Described in
another section.
Press this key.
Main operation flow
Operation flow
OLP3003
F4
F5
F5
Referring to object
analysis tests of the
calibration analysis
Selecting an
analysis test
(R) Routine
(T)Test Requisition
(C)Calibration
Referring to the
arrangement of
calibration samples
OLP1015E
ALARMCLEAR
?
STANDBY
Help
Exit
Cal.
Calibration
1997/09/03 10:10
Blue
Yellow
Green
Type:
Cuvette:
Profile:
*
: Individual selection
:RB Only
:One Point
:All
Calibration Options
ITM01 ITM02 ITM03 ITM04 ITM05 ITM06 ITM07 ITM08 ITM09 ITM10 ITM11
Start Entry
Individual
Requisition(R)
Analyzer
Requisition(C)
F4 (Start Entry)
Help
Exit
RB All
Change
Option
Select
All Test
Entry
ALARM CLEAR
Description:
Use this screen to requisition for reagent blanks and calibrators in racks. Tests with expired
calibrations, and tests with new bottles or lot numbers of reagent (depending on Advanced
Calibration programming in [Calibration Specific] ) are automatically requisitioned for
calibration after the reagent volume check.
Type
: The sample type can be changed to serum, urine, or other. Select the sample type from
the pull-down menu.
Cuvette
: Use this pull-down menu to select the inner wheel, the outer wheel or both.
Profile
: When function key F4 (start entry) is pressed, the profile field becomes active. This
field can be used to select an existing profile.
Analyzer Requisition:
Use this button to perform an automatic calibration requisition for tests
with expired cal or new bottles/lots using Advanced calibration.