Automated Chemistry Analyzer
AU2700 User's Guide
Chapter C - Basic Operations
C-49
May, 2002
Processing an Emergent Sample Using the STAT Table and a Sample Cup with
Barcode
If a sample cup with a barcode attached is used, care must be taken when setting it on the STAT table.
A sample cup with a barcode attached should be set on the STAT table with the barcode facing in the
direction as shown in the figure below.
OLU2115E
The sample cup should be placed
in an outer hole on the STAT table
with the barcode facing outward.
8
7
6
5
4
3
2
1
22
21
20
19
18
17
16
15
14
13
12
11
10
9
U-L
U-H
S-L
K-SEL-Na
S-H
CLEAN
The barcode is not read for
cups placed in the inner holes.
STAT table