D-182
Chapter D - Software
Automated Chemistry Analyzer
AU2700 User's Guide
March 31, 2005
5.3 Consumable Management
F3
OLP3064
(M)Maintenance
(C)Consumable Management
Outputting the
current state
of consumables
to a data list
ALARMCLEAR
?
STANDBY
Help
Exit
Change
Consumable Management
11/30/2000 10:10
OLP1cons
No.
Part
Changed
Used
1. Lamp
03/10/2000
148 2000 hours
2. S. Probe
3. R. Probe
4. S. Syringe
5. R. Syringe
6.
7.
8.
9.
10.
11.
12.
13.
14.
15.
16.
History
Entry
Delete
Description:
Use this screen to track replacement and supply periods of consumables. This list makes it easy
for maintenance personnel to schedule consumable purchasing and to observe maintenance
periods. When a part achieves its projected life expectancy, the background screen color for
that part changes to red. This color change does not indicate that the part must be replaced. To
determine when to replace the part, consult the specific part guidelines.
No.:
Provides a number for the part listed.
Part:
Displays the name of the consumable.
Changed:
Displays the date and time the consumable was changed.
Used:
Shows the consumption rate in days, weeks, months, or years. The average life span of
each consumable is also displayed.
Part No.:
Displays the part number of the consumable to be edited. Use the </> buttons to
select a different part number.
Part Name:
Type a part name using 20 characters or less.
Unit:
Used to set the method for calculating the consumption and average live of the
consumable. Select days, weeks, months, or years from the pull-down menu. Use the field that
follows the unit field to add the life expectancy.