C-120
Chapter C - Basic Operations
Automated Chemistry Analyzer
AU2700 User's Guide
May, 2002
7.
Operation Guide.
OLP4073E
Power ON
(Press the Power-ON switch.
)
Using the System Status key: See "D-14".
Using the [Calibration] test requisition menu: See "C-18, D-77".
Using the [QC] test requisition menu: See "C-20, D-73".
Check the repeat-run
work list.
↓
Perform requisition
operation for repeat run.
↓
Set the samples.
( STAT table useable
)
Using the [FD Data Management] menu: See "D-370".
Using the [Start Condition] menu: See "D-52".
Open the main valve to the deionized water.
On the startup screen,
set the index date and time,
round, and start No.
Check the remaining reagent.
Requisition operation for calibration analysis
Requisition operation for quality control analysis
Using a worksheet created by an external device, and using the
[Normal] test requisition menu: See "C-22, D-57".
Requisition operation for normal analysis
Use the start key.
Start analysis.
Set the samples.
Create a worksheet using AU2700.
Emergent sample measurement
Auto repeat run
Setck and edit the analysis data.
Close the main valve to the deionized water.
Perform repeat
run manually?
Normal analysis
requisition operation
↓
Set the sample.
↓
Press the STAT test switch
.
Change
round?
End (Press the End processing key.
)
Standby mode
Warmup mode
Standby mode
YES
NO
NO
YES
Initialize mode
Login
Perform inspections.
Create a report and master data file
.
Save analysis data.
Using the [Report] menu and [Data List] menu: See "D-175, D-178".
Details of inspection methods: See "F-6".
Set on the rack feeder.
Set on the STAT table.
Check the quality control data.
Check the quality control data.
Overwrite with repeat-run data.
Edit the data
Correct the data.
Using the [QC Monitor] menu: See "D-123, D-132".
Using the [Normal] test requisition menu: See "D-57".
Using the [Retrieve Repeat Data] menu: See "D-83".
Using the [Data Edit] menu: See "D-159".
Using the [Data Correction] menu: See "D-167".
Repeat run
Check for leaking sample dispenser or reagent dispenser.
Check the amount of master detergent.
Using the [Repeat]
test requisition
menu: See "D-67".
Using the [Repeat]
test requisition
menu: See "D-67".
Supply reagent.
The reagent should be prepared and set the previous day
or prepared and set at least one hour before starting analysis.
Perform inspections.
Details of inspection methods: See "F-6".
Check the sample probe, reagent probe and mixing bar.
Check the printer and paper.
Reagent tray
exchange or
reagent setting
change.
: Occasional operation
: Normal operation
: Mode change
: Supplement, reference
Using the System Status key: See "D-14".
Check the remaining reagent.
Calibrating ISE (option)
Cleaning ISE (option)