Automated Chemistry Analyzer
AU2700 User's Guide
Chapter C - Basic Operations
C-13
May, 2002
4. Calibrate the ISE:
A. Remove the small STAT table cover.
B. Press the STAT ROTATION/DIAG Button to rotate the STAT table so positions S-H,
S-L, U-H and U-L can be accessed.
C. Place the serum and/or urine high and low standards in the labeled positions on the
STAT table.
8
7
6
5
4
3
2
1
22
21
20
19
18
17
16
15
14
13
12
11
10
9
W2
W1
U-L
U-H
S-L
K-SEL-Na
S-H
CLEAN
8
7
6
5
4
3
2
1
22
21
20
19
18
17
16
15
14
13
12
11
10
9
U-L
U-H
S-L
K-SEL-Na
S-H
CLEAN
OLU2003
Top view of STAT table
STAT ROTATION/DIAG
switch
STAT table
Sample cup
STAT table cover
(small)
D. Replace the small STAT table cover.
E. Select [System Status], [ISE Status].