Automated Chemistry Analyzer
AU2700 User's Guide
Chapter D - Software
D-195
May, 2002
5.5.4 Alarm Log
(M)Maintenance
(M)Maker Maintenance
(L)Alarm Log
ALARMCLEAR
Help
Exit
Re-Display
Alarm Log
OLP1ala
Alarm Message:
Time:
1. 9004 MENU [Alarm Log]
11:42:51
2. 9005 MENU [Program Version] EXIT
11:42:00
3. 9004 MENU [Program Version]
11:41:02
4.
5.
6.
7.
8.
9.
10.
11.
12.
13.
14.
15.
16.
Period: 11
/28/2000
- 11/30/2000 Date: 11/30/2000
Select the function using the function key or the mouse.
>
?
STANDBY
11/30/2000 10:10
Description:
This screen displays an alarm list of previous alarms and screens accessed on the system. Use
this menu to determine recent or common alarms (past alarms).
If an alarm is occurring at this time, corrective actions are located in the alarm help function. To
start alarm help, select the icon located at the top of the screen with the triangle and exclamation
point.
Use this software screen to:
Display an alarm list:
1. Select [Maintenance], [Maker Maintenance], [Alarm Log].
2. Use the pull-down menu or < /> buttons to select a date on which you wish to view the
alarm log.
Print an alarm list:
1. From the [Alarm Log] screen press function key F3 (print)
2. Choose "Displayed Date" and "All Date." Displayed date only prints the date selected.
"All date" prints alarm logs for all dates.