Automated Chemistry Analyzer
AU2700 User's Guide
Chapter D - Software
D-65
May, 2002
2.6.1 Daily Control
F6
F7
F3
F4
F8
F4
OLP3029
F6
Specifying a control
number and an
analysis test
Referring to a daily
variation chart of the
specified control
number
(R) Routine
(Q)QC Monitor
(D)Daily Control
Displaying the past
5 times of data
status on each
control samples
Referring to the
daily variation data
Outputting the daily
variation data in a
data list
Registering a
desired comment for
the displayed daily
variation data
Setting the display
conditions
One of the tabs
can be selected
Button function
Described in
another section.
Press this key.
Main operation flow
Operation flow
OLP1021
QC:
Help
Exit
Control
No.
Qnnn Qnnn (Qnnn - Qnnn)
Sample No.
Index:
Control Name:
yyyy / mm / dd hh : mm
(ALL)
Serum
Type:
Daily
Data
Comment
Change
Test
Select
QC Monitor
ALARMCLEAR
?
STANDBY
Daily Control
1997/09/03 10:10
ITM01 ITM02 ITM03 ITM04 ITM05 ITM06 ITM07 ITM08 ITM09 ITM10 ITM11
Description:
Use this screen to observe quality control data on a specified analysis date. Also, QC daily
statistics and graphs can be viewed, comments can be added, and a list of QC data can be
printed. To compare the average of quality control data, go to the day to day control screen.
Index
: Select the index (date/time) from the pull-down menu.
Type
: The sample type can be changed to serum, urine, or other. Select the sample type from
the pull-down menu.
Control Name
: Select a control number and name from the pull-down menu.
Test - List:
Use the mouse to deselect and select tests.
QC Sample No
.: Enter a sample number within the specified range. Enter an asterisk (*) to
select all samples. Enter a space if no selection is required.
"QC Monitor" Tab Screen:
Index
: Select the index (date/time) from the pull-down menu.
Type
: The sample type can be changed to serum, urine, or other. Select the sample type from
the pull-down menu.