Automated Chemistry Analyzer
AU2700 User's Guide
Chapter C - Basic Operations
C-51
May, 2002
Flex Rack: Adapter Installation
1. Match the adapter slit and the rack slit as shown in the diagram below.
2. Slide the adapter guide into the guide groove of the rack and push the adapter into the
rack.
3. A clicking sound verifies that the adapter lock is engaged on the rack.
1
2
3
4
5
6
7
8
9
10
Adapter lock
Guide
Adapter
Slit
Push in
Rack window
Guide groove
Rack
Slit
Flex Rack: Adapter Removal
1. With your finger, push the adapter lock lightly from the outside of the rack window to
disengage the lock.
2. When the upper edge of the adapter comes out from the rack, grasp the adapter and pull it
out.
Adapter lock
Rack window
Push up lightly
with a finger