C-14
Chapter C - Basic Operations
Automated Chemistry Analyzer
AU2700 User's Guide
May, 2002
OLP1014
ALARMCLEAR
?
1997/09/03 10:10
STANDBY
Slope
Date/Time
ISE Stop
ISE Unit Start
Prime
Type Change
Urine
Reagent
Ready
Electrode
MID Solution Factor
Na
K
Cl
60.0
61.1
59.0
0.0
56.4
55.3
52.4
52.2
55.1
54.3
54.7
0.0
51.2
49.5
55.4
55.2
–49.0
–52.7
–50.0
0.0
–48.9
–59.2
–52.3
–51.0
1.005
1.006
1.005
1.016
1.003
1.111
1.026
0.999
1.013
0.997
1.000
0.986
1.003
9.999
1.259
–7.945
1.001
1.000
0.998
0.999
1.000
3.001
0.222
1.003
Na
K
Cl
ISE Status
Normal Range
38-68
38-68
-38 - -68
0.80-1.20
0.75-1.25
0.75-1.25
Help
Exit
Result Cal.
Result Select.
Slope Chart
1994 / 12 / 18
1994 / 12 / 18
1994 / 12 / 18
1994 / 12 / 18
1994 / 12 / 18
1994 / 12 / 18
1994 / 12 / 18
1994 / 12 / 18
13 : 33
12 : 04
11 : 56
9 : 00
18 : 44
13 : 01
18 : 42
12 : 33
F. Select function key F5 (ISE unit start) Highlight: A. Cal for Serum, B. Cal for Urine, or
C. Cal for Serum and Urine. When the prompt “Start Operation” appears, select “Yes.”
It takes approximately 4 minutes to calibrate for serum and urine.
Caution
After analysis is started, keep your hands away from moving parts.
G. When the calibration is complete, the results are displayed in a list. Verify the slope and
mid-factor values are in range. If the values are out-of-range, (highlighted in yellow)
perform calibration again. If the slope value or mid factor is still out of the normal
range, the electrode may be deteriorating.
Slope
38 to 68 (Na and K are positive. Cl is negative)
Mid Factor
Na = 0.80 to 1.20
K = 0.75 to 1.25
Cl = 0.75 to 1.25
H. After checking results, press function key F2 (exit).
The main screen returns.
See Also
For calibration errors, refer to the ISE section of the Troubleshooting Chapter. For information about
replacing electrodes, refer to the Maintenance Chapter.
Содержание AU2700
Страница 17: ...A 4b Automated Chemistry Analyzer AU2700 User s Guide November 1 2009 This page intentionally left blank ...
Страница 178: ...D 4 Chapter D Software Automated Chemistry Analyzer AU2700 User s Guide March 31 2005 ...
Страница 260: ...D 86 Chapter D Software Automated Chemistry Analyzer AU2700 User s Guide May 2002 ...