Automated Chemistry Analyzer
AU2700 User's Guide
Chapter C - Basic Operations
C-101
May, 2002
Processing a Priority Sample using the STAT Table
Important
The 22 outer positions on the STAT table are assigned to STAT, repeat, reagent blank, calibration, or QC
functions in the [System Maintenance] menu. Verify which positions can be used for serum, urine, or
other priority STAT analysis on the analyzer.
1. From the main menu select [Routine], [Test Requisition], and [Normal], in that order.
The
screen for requisitioning normal analysis is displayed.
2. From the “S. No.” field, type “P,” then enter.
The next available priority sample number
will display.
3. From the “Type” field, use the pull-down menu to select serum, urine, or other.
4. Requisition tests individually, by profile, or batch. If requisitions are required for more
than one sample type, select function key F2 (exit) to return to “Type.”
5. Press the STAT button to turn the STAT SET LED on. The STAT SET LED must be on
before STATS can be loaded. Do not load samples if the SET LED blinks or does not
come on. The SET LED turns on when loading is possible. If necessary, press the STAT
ROTATION/DIAG Button to rotate to the appropriate position to load samples.
6. Place the sample in a position assigned for the sample type on the table.
If STATS are not
barcoded, they must always be placed in increasing numerical order on the table! Always
remove STATS from the table as soon as the END LED is on.
Warning
Close the STAT table cover after placing samples. A consistent temperature in the STAT table is
important to the accuracy of the results.
7. Press the STAT Button.
Analysis starts.
8. The END LED comes on when STAT sample aspiration is complete. Remove containers
from the STAT table.