D-76
Chapter D - Software
Automated Chemistry Analyzer
AU2700 User's Guide
February 4, 2004
Use this software screen to:
Edit Results:
For procedures, refer to "Editing Results" in the Basic Operations Chapter.
Edit flags:
Flags on the sample results can be modified.
1. From the main menu select [Routine], [Data Management], [Data Edit].
2. Select a sample type and sample number or ID.
3. Press function key F5 (Data Display).
4. Click function key F5 (Edit Result).
A window appears.
5. Use your mouse to click on the "Flag" field. Now enter a flag.
6. After the entry is completed, select "Close."
The settings are saved.
7. Press function key F2 (exit) twice.
Add a search key:
In addition to a sample type and analysis test, patient information, such as sex, and age can be
specified as a search key for data search.
1. From the main menu select [Routine], [Data Management], [Data Edit].
2. Press function key F6 (search add).
The “Adding a search key” screen is displayed.
3. Set “Sex,” “Age (Low and high limit)," and 6 types of demographics.
4. After the entry is completed, select “YES.”
The settings will be saved as search keys.
Edit a sample in real-time mode:
For procedures, refer to "Editing Results" in the Basic Operations Chapter.
Edit a sample ID in batch or no host mode:
For procedures, refer to "Editing Results" in the Basic Operations Chapter.
Precaution:
If your analyzer is being operated in Sequential (non-barcoded) Mode and data editing is conducted
during the analyzer’s MEASURE modes (analyzer modes other than STOP, PAUSE, STANDBY or WARM
UP) a database conflict may occur. If a database conflict occurs, it will result in a sample concordance
error regarding the identification of a specific sample and the test results associated with the sample.
If you operate, or begin to operate, your analyzer in
Sequential Mode
, you must cease all Data Edit
activities until
AFTER
the analyzer has shifted to the STOP, PAUSE, STANDBY, or WARM UP mode.
Do NOT edit
any data during analyzer
MEASURE modes
(MEASURE 1, MEASURE 2, MEASURE3
OR MEASURE END) as indicated on the analyzer screen mode display area.