Automated Chemistry Analyzer
AU2700 User's Guide
Chapter D - Software
D-15
May, 2002
1.4 ISE Status
Referring to the data
for samples under
analysis
F5
Referring to main
information on
system status
[System Status]
(4) Data
Display
Referring to the data
for all samples
Referring to the data
for ISE (option)
Referring to the data
for samples analyzed
with the STAT table
Select [All Data] tab
"Data Display" screen
Select [STAT Data] tab
Select [ISE Data] tab
F6
F5
Referring to data
related to analyzed
smples
F5
OLU3013E
One of the tabs
can be selected
Button function
Described in
another section.
Press this key.
Main operation flow
Operation flow
OLP1014
ALARMCLEAR
?
1997/09/03 10:10
STANDBY
Slope
Date/Time
ISE Stop
ISE Unit Start
Prime
Type Change
Urine
Reagent
Ready
Electrode
MID Solution Factor
Na
K
Cl
60.0
61.1
59.0
0.0
56.4
55.3
52.4
52.2
55.1
54.3
54.7
0.0
51.2
49.5
55.4
55.2
–49.0
–52.7
–50.0
0.0
–48.9
–59.2
–52.3
–51.0
1.005
1.006
1.005
1.016
1.003
1.111
1.026
0.999
1.013
0.997
1.000
0.986
1.003
9.999
1.259
–7.945
1.001
1.000
0.998
0.999
1.000
3.001
0.222
1.003
Na
K
Cl
ISE Status
Normal Range
38-68
38-68
-38 - -68
0.80-1.20
0.75-1.25
0.75-1.25
Help
Exit
Result Cal.
Result Select.
Slope Chart
1994 / 12 / 18
1994 / 12 / 18
1994 / 12 / 18
1994 / 12 / 18
1994 / 12 / 18
1994 / 12 / 18
1994 / 12 / 18
1994 / 12 / 18
13 : 33
12 : 04
11 : 56
9 : 00
18 : 44
13 : 01
18 : 42
12 : 33
Description
Use this screen to display the data and graph of the slope of the calibration curves for Na, K,
and Cl. It is also possible to display the mid-standard solution factor and selectivity check
results.
See Also
For information on correcting many of the following abnormal messages, refer to the “Fatal and Nonfatal
Errors” section located in the Flags Chapter.
Use this software screen to:
Check ISE Status:
Ready/Busy/Stop/No Connection
Indicates the operation mode of the ISE.
M
essage
M
eaning
C
hange
in
baCkground
Color
READY
Ready to start analysis
Blue
BUSY
Analysis in progress
Red
STOP
ISE in stop mode
Yellow
Unconnected
ISE is disconnected from system
(Communication error or power failure)
Yellow
Electrode
Electrode indicates whether the latest slope value falls within the range of normal values.
M
essage
M
eaning
C
hange
in
baCkground
Color
Electrode
Slope values in range
Blue
Electrode
Slope out of range
Yellow
Содержание AU2700
Страница 17: ...A 4b Automated Chemistry Analyzer AU2700 User s Guide November 1 2009 This page intentionally left blank ...
Страница 178: ...D 4 Chapter D Software Automated Chemistry Analyzer AU2700 User s Guide March 31 2005 ...
Страница 260: ...D 86 Chapter D Software Automated Chemistry Analyzer AU2700 User s Guide May 2002 ...