Automated Chemistry Analyzer
AU2700 User's Guide
Chapter D - Software
D-79
May, 2002
F8 : Test No.
Where the data is to be searched by the test item, it can be specified by the test No.
1. Press the F8 (Test No.) key. The “Select Test No.” dialog will appear.
2. Enter a test No. (excluding No.96 and test numbers specified as an inter-item calculation
test) of analysis tests to be referred to.
3. Select the “OK” button. The specified test name will be displayed.
Tip
It is also possible to specify another test item from the “Test Name” pull-down list.
Use this software screen to:
Correct results:
For procedures, refer to "Correcting Results" in the Basic Operations Chapter.
Recalculate data:
Concentration values for sample data can be recalculated based on the acceptable cal. OD values.
Do this only if cal. process is set to OD in the [Cal. Specific] screen.
As a result, the existing
analysis data will be overwritten by the data after recalculation. Also, analysis tests subject to
recalculation can’t be printed for a report. Data can be sent online. A "c" flag is attached.
Important
Data cannot be recalculated if changes are made to the calibration parameters. For example if a failure
was due to the wrong set point, or range, the calibration must be run again.
1. From the [Data Correction] screen, select from serum, urine, other; and routine,
emergency, and STAT.
2. Press function key F6 (recal. data).
A window to execute recalculation is displayed.
3. Select “OK.”
Data recalculation starts.
Tip
Data saved as an OD value can be modified by recalculating. Data saved as a concentration value cannot
be modified.