Measurements
R&S
®
RTO6
361
User Manual 1801.6687.02 ─ 05
If "Continuous auto scale" is enabled, the instrument determines the number of bins
based on the timebase, the current measurements, and other settings. To set the num-
ber of bins manually, disable "Continuous auto scale".
Remote command:
MEASurement<m>:STATistics:HBINs
Enable (Track)
Enables the track of measurement results over time and displays the track waveform. It
is the graphical interpretation of all measurement values of a single acquisition.
The track is available for most amplitude/time measurements (except for High, Low,
Amplitude, Max, Min, Peak to peak, Mean, RMS, S-dev, Pos. and Neg. overshoot, and
Area), and for jitter measurements.
Enabling the track enables also the
and
Before you can enable the track, activate the appropriate measurement.
Remote command:
Reset
Immediately resets the histogram, the long-term measurement and the statistics.
Stopping and starting the acquisition does not reset these analyses but only stops and
continues them.
To delete all results, waveforms and history, select "Display" menu > "Clear all".
Remote command:
MEASurement<m>:STATistics:RESet
Analysis scale
The measurement scale of a long-term measurement diagram or measurement histo-
gram can be set automatically by the instrument, or manually.
Use automatic scaling if the measurement is running and you cannot see the expected
results.
"Continuous auto scale"
Performs an automatic scaling whenever the long-term waveform or
the histogram does not fit in the diagram during the measurement
period.
"Vertical"
Defines the vertical scaling per division for long-term measurement
period and the measurement histogram.
"Offset"
Defines an offset for the long-term measurement and the measure-
ment histogram.
Remote command:
MEASurement<m>:VERTical:OFFSet
Automatic measurements