Remote control commands
R&S
®
RTO6
1798
User Manual 1801.6687.02 ─ 05
Parameters:
<Bitrate>
Range:
100 to 5E+6
Increment: 1
*RST:
100E+3
Default unit: bps
BUS<m>:CAN:FDATa:ENABle
<CANFDEnabled>
Enables the CAN FD protocol configuration.
The setting is available in CAN FD option.
Suffix:
<m>
.
1..4
Parameters:
<CANFDEnabled>
ON | OFF
*RST:
ON
BUS<m>:CAN:FDATa:DBITrate
<FlexDatBitrate>
Sets the bit rate of the data phase. The data rate can be higher than the arbitration
rate, but it is uniform and fixed for a given CAN FD bus.
The setting is available in CAN FD option.
Suffix:
<m>
.
1..4
Parameters:
<FlexDatBitrate>
Range:
100 to 15E+6
Increment: 1
*RST:
1E+6
Default unit: bps
BUS<m>:CAN:FDATa:SAMPlepoint
<FlexDatSmpPt>
BUS<m>:CAN:SAMPlepoint
<SamplePoint>
Sets the position of the sample point within the bit in percent of the nominal bit time.
Alternatively, you can set the sample point with
and
.
For CAN FD signals,
BUS<m>:CAN:SAMPlepoint
defines the synchronization of the
arbitration phase, and
BUS<m>:CAN:FDATa:SAMPlepoint
defines the synchroniza-
tion of the data phase.
Suffix:
<m>
.
1..4
Selects the serial bus.
Protocols