![R&S RTO6 User Manual Download Page 1714](http://html2.mh-extra.com/html/rands/rto6/rto6_user-manual_30646461714.webp)
Remote control commands
R&S
®
RTO6
1714
User Manual 1801.6687.02 ─ 05
TRIGger<m>:SPI:DPOPerator
<DataPosOperator>
Sets the operator for the data position. You can defined an exact position, or a position
range.
Parameters:
<DataPosOperator> ANY | OFF | EQUal | GETHan | INRange | RANGe
ANY = OFF
The position of the required pattern is not relevant.
EQUal | GETHan
Equal, Greater or equal than. These conditions require one data
position to be set with
.
INRange = RANGe
Set the minimum and maximum value of the range with
.
*RST:
ANY
TRIGger<m>:SPI:DPOSition
<DataPosition>
Sets the number of bits or words to be ignored before the first bit or word od interest.
The effect is defined by
.
Parameters:
<DataPosition>
The index 0 is associated with the first data byte.
Range:
0 to 4095 for triggering on one line (MISO or
MOSI), 2047 for triggering on both lines.
Increment: 1
*RST:
0
TRIGger<m>:SPI:DPTO
<DataPositionTo>
Defines the last bit or word of interest, if
INRange
.
Parameters:
<DataPositionTo>
Range:
1 to 4095 for triggering on one line (MISO or
MOSI), 2047 for triggering on both lines.
Increment: 1
*RST:
1
TRIGger<m>:SPI:FCONdition
<DataOperator>
Selects the operator for the MISO and MOSI pattern.
Parameters:
<DataOperator>
EQUal | NEQual
*RST:
EQUal
Protocols